Proceedings of 2015 Photonics & Electromagnetics Research Symposium, Prague, July 6 - 9,Page(s)2746-2749
Abstract
In this paper measurements of temperature increase of three different hip replace-
ments during MRI were carried out. The hip replacements were embedded in an agar phantom
which mimics dielectric properties of muscle tissue and were placed into a MRI system with the
static magnetic field intensity of B0 = 1.5 T. The MRI of the samples took in total 20 minutes.
Thermograms of the hip replacements were taken with a thermographic camera in two different
time instances.
Citation
Jan Vrba,
and
Miroslav Wiewegh,
"Measurement of Temperature Increase of Metal Hip Replacements During Magnetic Resonance Imaging," Proceedings of 2015 Photonics & Electromagnetics Research Symposium, Prague, July 6 - 9,Page(s)2746-2749